Clinical Computer Vision for Neonatal Jaundice Severity Estimation Under Variable Skin Tone Calibration: A Fairness-Audit Evaluation of Transcutaneous Bilirubin Threshold Safety
By Ifeoma Chidinma Okonkwo and Aisha Bello Usman.
Abstract
Clinical computer vision systems that estimate neonatal jaundice severity from skin images must operate across a wide range of skin tones, yet the safety of the transcutaneous bilirubin thresholds they inform has not been characterised under realistic calibration error. The framework is formalised as a threshold-safety model in which calibration bias shifts the effective decision boundary, and the resulting false-negative and false-positive rates are expressed as functions of bias magnitude, threshold margin, and the population distribution of bilirubin values. A replication-based evaluation of the before-after-control-impact estimator was conducted to characterise the design's operating characteristics under a declared variance structure (between-cluster SD = 6.0, residual SD = 4.5, intraclass correlation = 0.64), with power estimated across 2000 design cells. A calibration bias of 2 mg/dL can eliminate the safety margin of a 15 mg/dL referral threshold for a substantial fraction of the neonatal population, and that the bias required to produce a given false-negative inflation depends jointly on the threshold margin and the local prevalence of severe hyperbilirubinaemia. The framework provides a structured method for auditing threshold safety before clinical deployment and identifies the conditions under which skin-tone calibration error becomes clinically consequential.