Surgeries
Comparison of Two Fluid-Management Platforms for Flexible Ureteroscopy in Renal Stone Treatment: Pressure-Feedback Irrigation and Aspiration Versus a Suction Access Sheath Without Pressure Feedback
By Petrisor Aurelian Geavlete, Cristian Mareș, Cătălin Andrei Bulai and 4 others.
Abstract
Background/Objectives: In the modern treatment of renal stones, intrarenal pressure management is an important safety factor in flexible ureteroscopy. Modern techniques that include real-time monitoring of intrarenal pressure, continuous irrigation, and controlled aspiration can improve intraoperative visibility while reducing the risk of problems associated with increased pressure.
Methods
A retrospective comparative observational study was performed on 120 consecutive eligible patients with single pyelocaliceal renal stones and a negative preoperative urine culture. Group 1 included 55 patients treated with the ClearPetra® 10/12 Fr. suction access sheath, without intrarenal pressure monitoring. Group 2 included 65 patients treated with the Elephant IV 10/12 Fr. sheath associated with the Tidor™ device, which allows intrarenal pressure monitoring and automatic control of irrigation and aspiration. In Group 2, the intrarenal pressure limit was set at 30 mmHg. Access sheath insertion was possible in all cases.