WFUMB-Recommended Versus Manufacturer-Optimized Protocols for Ultrasound-Guided Attenuation Parameter in Hepatic Steatosis Assessment: A Retrospective Single-Vendor Multicenter Study
By Hidekatsu Kuroda, Takuma Oguri, Naohisa Kamiyama and 16 others.
Abstract
Background/Objectives: The ultrasound-guided attenuation parameter (UGAP) is conventionally measured at a 4 cm start depth from the skin with a 4 cm region of interest (ROI), whereas 2024 World Federation for Ultrasound in Medicine and Biology (WFUMB) guidance recommends 2 cm from the liver capsule with a 3 cm ROI. We compared these two ROI configurations, applied to the same archived acquisitions, against MRI-derived proton density fat fraction (MRI-PDFF).
Methods
This retrospective study included 565 patients with chronic liver disease from four of six participating Japanese centers at which raw data had been stored. Conventional UGAP was measured in real time; WFUMB-based values were calculated retrospectively from the same acquisitions, without real-time operator optimization. Performance was assessed by receiver operating characteristic analysis and the Obuchowski index, with integrated discrimination improvement (IDI) and net reclassification improvement (NRI) as secondary analyses.