Journal of applied clinical medical physics
Automated offline adaptive radiotherapy through CBCT imaging: A multidimensional feasibility study of hypersight and iterative CBCT
Par Archibald-Heeren BR, Leong A, Middlebrook N et 5 autres.
Contenu en anglais
Résumé
Background
Offline adaptive radiotherapy (OffART) offers potential dosimetric and clinical advantages by adjusting planned dosimetry in response to systematic patient changes throughout treatment, however implementation is associated with an increased labor burden. Technological innovations in automated contouring, image registration and on-board imaging present the opportunity for an automated workflow that achieves the benefit of OffART with a reduced impact on staff workload.



